RELEASE PEDAL ADJUSTMENT

English
Tools Required:
• None
Procedure:
1 .
Manually disengage the release pedal swivel from the release pedal assembly .
2 .
To increase the release rod engagement with the release valve, turn the release pedal swivel clockwise on the
threaded release rod .
3 .
To decrease the release rod engagement with the release valve, turn the release pedal swivel counterclockwise
on the threaded release rod .
Note: If the pedal swivel assembly is threaded too far onto the release rod, the release valve will be partially activated
and the jack will drift .
Return To Table of Contents