Herunterladen Inhalt Inhalt Diese Seite drucken

Pneumatic Fowler Adjustment - Stryker 747 Wartungshandbuch

Transport-stretcher
Inhaltsverzeichnis

Werbung

PNEUMATIC FOWLER ADJUSTMENT

English
Tools Required:
• 5/32" Hex Allen Wrench
• 1/2" Combination Wrench
Procedure:
1 .
For easier access, raise the Fowler to 75° or higher .
2 .
Using a 1/2" combination wrench, loosen the hex nuts in the actuator arms on the end of the trip bar (item M) .
3 .
To adjust the Fowler, use a 5/32" hex Allen wrench to turn the Allen screws 1 to 2 turns counterclockwise if the
Fowler will not move or 1 to 2 turns clockwise if the Fowler will not hold its position .
4 .
Retighten the hex nuts . Be sure the Fowler travels from flat up to 90° and down again and holds its position when
weight is applied before returning the stretcher to service .
Return To Table of Contents
1-28
Service Information
0747-009-006 REV J
www.stryker.com