Herunterladen Inhalt Inhalt Diese Seite drucken

Description; Product Overview; Intended Use; Limitations On Use - Dräger X-am 3500 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für X-am 3500:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 5
Abbreviation
Explanation
STEL
Short time exposure limit, threshold of an
exposition over a short period of time (gen-
erally 15 minutes).
TWA
Time weighted average, average shift val-
ues are generally limited to eight hours
exposure per day per workplace for 5 days
a week during a work lifetime. Observe the
national definition of the occupational
exposure limit.
3

Description

3.1

Product overview

The graphics are displayed on the fold-out page.
Graphic A
1 Display
2 Locking screw for an addi-
tional charging module
3 Power supply unit
Graphic B
1 Gas inlets
2 Thread port for pumps
and calibration adapter
Graphic C
1 LED green/yellow/red
Graphic D
1 feature button 1
2 Feature button 2
Graphic E
1 Status information
Graphic F
1 Clip (optional)
Graphic G
1 Alarm A1, continuous red
light
2 Alarm A2, alternating
red/black light
Graphic J
1 Battery pre-alarm
Graphic K
1 Calibration adapter (grey
coloured ring)
Instructions for use
|
X-am
4 Charge LED green/red
5 Induction charger
3 Horn
4 Pump outlet and inlet
2 LED yellow/red
3 Feature button 3
2 Navigation bar
3 STEL alarm
4 TWA alarm
2 Battery main alarm
3 Exhaust
®
3500
Graphic K
2 Gas inlet
Graphic M
1 Pump adapter (blue
coloured ring)
2 Gas inlet
3 Dust and water filter
3.2

Intended use

®
Dräger X-am
3500 is a portable gas detector for clearance
measurements and for the continuous monitoring of the
concentration of several gases in ambient air in the workplace
and in explosion-hazard areas.
The X-am 3500 allows measuring up to 4 gases in
accordance with the DrägerSensors (XXS O
XXS CO LC, XXS NO
, XXS SO
2
are installed. The gas detector can be operated in pump mode
or in diffusion mode.
3.3

Limitations on use

The gas detector is not suitable for measuring process
gasses.
Operation of the gas detector in the charging cradle in a
vehicle is only permitted under the following conditions:
– Acoustic signalling must be deactivated (with the help of
the PC software CC-Vision) so that the driver is not
distracted.
3.4
Approvals
A copy of the rating plate, the declaration of conformity and
the sensor data relevant for measurement purposes are
provided in the enclosed supplementary documentation (part
no. 90 33 655).
Explosion protection:
BVS 17 ATEX E 040 X certifies the intended use in explosion-
hazard areas and the measuring function for explosion
protection. PFG 19 G 001 X certifies the measurement of
oxygen deficiency and oxygen surplus as well as the
measuring function for toxic gases. See the supplementary
documentation for certified gases and measured values (part
no. 90 33 655).
CSA-specific information:
Only the combustible gas part of this instrument has been
tested for measuring accuracy.
Radio approval (X-am 8000 only):
The information for radio approval can be viewed in the menu.
For further information, see the following chapter: "Opening
information", page 32.
|
Description
4 Locking screw
4 Exhaust
5 Locking screw
, XXS H
S LC,
2
2
and CatEx 125 PR) which
2
en
27

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis