Preparation for Use
4 – 5 Installing the irrigation tube
1
OPEN
2
● Make sure that the tube is securely set on the rollers when closing the pump cover� If the tube is not correctly
positioned on the rollers and the cover is closed, the tube could be cut or sheared�
● Do not use non-NSK genuine irrigation tubing�
● When mounting the irrigation tube, be careful not to get your hand caught in the moving part of the irrigation
pump�
● Do not mount the irrigation tube while the power is ON�
● Do not use the irrigation tube if the pouch is punctured or the seal is broken� It is not sterile�
● Federal law restricts this device to sale by or on the order of a physician�
"Needle"
side
CAUTION
12
3
CLOSE

Quicklinks ausblenden: