Herunterladen Inhalt Inhalt Diese Seite drucken

NSK Surgery System Surgic Pro2 Gebrauchsanweisung Seite 35

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 55
Setting
Main menu
Sub menu
FC SET
FC MODE
SLEEP
● When setting values are changed, press the enter key to finalize� If you press the SPEED key, TORQUE key, or Cancel
key without finalizing with the Enter key or if you turn the power off, setting details will not be applied�
● Initializing settings returns "UNT SET" in setting mode and program settings to factory default� "BLE SET" and "FC
SET" in setting mode are not initialized� Make a memo or other record of the current settings as necessary�
● Finalized setting details will be retained even if the power is turned off�
Setting details
AUT: Set the foot control to the automatic
connection�
FC: Enable sleep mode� (It may be
necessary to keep pressing the speed
control pedal of the foot control to connect
the control unit and the foot control
automatically�)
(See "5 – 4 Sleep mode")
Set the time until the foot control shifts to
sleep mode�
*This can be set only when FC MODE is
changed from AUT to FC�
(See "5 – 4 Sleep mode")
CAUTION
NOTICE
Setting range
AUT, FC
OFF
5m
10m
15m
20m
30m
45m
60m
34
Factory setting
AUT
5m

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis