7.4.4

Step 4: Setting the torque

Once the speed has been selected, the torque can be determined from the corresponding torque range.
Press the "Torque/Frequency" keys "▲" to increase or "▼" to decrease the maximum torque. By pressing the
key continuously the torques are shown in fast forward mode.
The following handpieces run on a fixed torque.
● K-Wire (Kirschner)
7.4.5

Step 5: Setting the pump supply quantity

Press the "Pump" keys "▲" to increase or "▼" to decrease the pump supply quantity. By pressing the key
continuously the values are fast forwarded.
To activate or deactivate the pump, press both "Pump" keys at the same time,
"Pump ▲ + ▼", or press "PUMP" on the pedal.
7.4.6

Step 6: Saving individual settings

By pressing the "Program"-key, 4 different settings can be saved for further use.
1.
Select the desired memory slot by pressing the „Program"-key (e.g. Program 1).
2.
Perform step 1 to step 5 for setting up the device.
3.
Repeat this procedure until all the memory slots are programmed.
4. At switch-off, the settings made by the user are automatically saved.
For the micro saws only the amount of coolant liquid can be altered.
By pressing both „HP + PR"-keys at the same time the memory slots can be reset to the factory default
settings.
OFA-Drill operating instructions, ref. 31595, V49/15
The build up of torque is displayed
as a bar graph. When maximum
torque is reached all elements of
the bar graph are full displayed.
● all micro saws