Herunterladen Inhalt Inhalt Diese Seite drucken

Step 4: Setting The Torque; Step 5: Setting The Pump Supply Quantity - Nouvag HighSurg 11 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
HighSurg 11
7.4.4

Step 4: Setting the torque

Once the speed has been selected, the torque can be determined from the corresponding torque range.
Press the "Torque/Frequency" keys "▲" to increase or "▼" to decrease the maximum torque. By pressing the
key continuously the torques are shown in fast forward mode.
When using the Shaver Handpiece in oscilating mode, the number of oscillations per second (number of
changing of directions per second of the rotating blade) of the Shaver blade is set. The direction arrow
shows both directions at the same time.
The following handpieces run on a fixed torque.
● Cont. Shaver
7.4.5

Step 5: Setting the pump supply quantity

Press the "Pump" keys "▲" to increase or "▼" to decrease the pump supply quantity. By pressing the key
continuously the values are fast forwarded.
To activate or deactivate the pump, press both "Pump" keys at the same time,
"Pump ▲ + ▼", or press "PUMP" on the pedal.
HighSurg 11, Operating Instructions, Ref. 31998, V14/15
The build up of torque is displayed
as a bar graph. When maximum
torque is reached all elements of
the bar graph are full displayed.
● Osc. Shaver
EN
13

Werbung

Inhaltsverzeichnis
loading