7. Maintenance, cleaning and recalibration
7. Lie the measuring instrument on the display side.
8. Screw the measuring instrument back together, starting with the individual screw on
the bottom.

7.2 Cleaning

1. Prior to cleaning, isolate the instrument properly from the pressure source and switch
it off.
2. Use the requisite protective equipment.
3. Clean the instrument with a moist cloth.
Electrical connections must not come into contact with moisture!
4. Wash or clean the instrument, in order to protect personnel and the environment from
exposure to residual media.

7.3 Recalibration

DKD/DAkkS certificate - official certificates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with
time intervals of approx. 12 months. The basic settings will be corrected if necessary.
WIKA operating instructions, model CPH6300
Downloaded from
www.Manualslib.com
CAUTION!
Physical injuries and damage to property and the environment
Improper cleaning may lead to physical injuries and damage to property and
the environment. Residual media in the dismounted instrument can result in
a risk to persons, the environment and equipment.
Wear the requisite protective equipment.
Carry out the cleaning process as described below.
CAUTION!
Damage to the instrument
Improper cleaning may lead to damage to the instrument!
Do not use any aggressive cleaning agents.
Do not use any hard or pointed objects for cleaning.