5. Commissioning, operation
Menu
"MENU"
SEt
CALC
1) Menu cannot be selected if there is data in the logger memory. If this is to be changed, the data must first be deleted (see chapter
5.5 "Data logger").
2) This menu can only be selected if an appropriate sensor is connected to connection 1. When using a second corresponding
sensor on connection 2 then the settings are adopted.
3) Only for 2-channel version
WIKA operating instructions, model CPH6300
Downloaded from
www.Manualslib.com
Parameters
Values
▲ or ▼
Auto power-off: Automatic switching off of the instrument
P.oFF
1 ... 120
oFF
Backlighting (factory setting: 5 s)
Lite
oFF
5 ... 120
on
Set calculation: Calculated displays (derived from sensor 1)
Selection of the calculation function
CALC
oFF
dP:dt
SPEd
FLo
"dP:dt": Settings
Time unit for the rate of change of the pressure
base
PEr.S
PEr.n
PEr.h
Test duration, for which the pressure change display is
t.int
calculated
00:01 ... 1:00
manuals search engine
Meaning
Auto power-off delay in minutes. If no button is
pressed and there is no data transfer via the
serial interface, the instrument will switch itself off
after this interval.
Auto power-off function inactive (continuous
operation)
No illumination
Illumination automatically switches off after
5 ... 120 s.
Illumination does not switch off
No calculated display
Rate of change of pressure
Airspeed through the orifice plate/pitot tube
Flow / Air flow
Change in pressure per second, e.g. mbar/s
Change in pressure per minute, e.g. mbar/min
Change in pressure per hour, e.g. mbar/h
Duration in [Minutes:Seconds], Factory setting:
0:01 = 1 s