5. Commissioning, operation
5. Using the [SET/MENU] button,
The display also gives an indication of the test result during the execution of the
test.
6. After 5 minutes, release the pressure.
7. Stop the logger.
Evaluation:
Read the data logger with GSOFT and enter "Commentary" in the register.
Originator
Test site
Date and time
Address / channel number
Inventory data of the object (channel type, nominal width, etc.)
Test procedures
Pressures, tolerances, etc.
Results of the testing
For this, a corresponding comments template can be prepared as a file. During printing,
company logos, etc. can be included.

5.5 Data logger

The instrument has two different logger functions:
Fvnc stoR
Fvnc CYCL
1-channel version:
The data logger displays 3 measuring results for each data set:
CALC off
Max.-Peak
CALC
Estimated calculations and dummy values (see chapter 5.4.5 "Rate of change of
pressure (only for sensor connection 1)")
WIKA operating instructions, model CPH6300
Downloaded from
www.Manualslib.com
Setting
t.int
: If very high values are set here, the display value can easily
be misinterpreted since the display values react correspondingly slowly!
The display value is calculated: ( P(T
If the data is not yet available over the entire
are extrapolated using the data already available.
Manual recording of the measured value by pressing the [STORE/
QUIT] button. Additionally, a measuring point input (L-Id) is required
Automatic recording at the interval of the set cycle time.
: Measured or mean value (depending on function), Min.-Peak and
"active": Measured or mean value (depending on function)
manuals search engine
dP.dt
can be changed on the display.
) – P(T
current
-
t.int
) ) /
current
t.int
timespan, the values
EN
t.int
23