Herunterladen Inhalt Inhalt Diese Seite drucken

Special Functions; Different Pressure Units; Different Measuring Modes - WIKA CPH6300 Betriebsanleitung

Hand-held druckmessgerät
Vorschau ausblenden Andere Handbücher für CPH6300:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5. Commissioning, operation

5.4 Special functions

5.4.1 Different pressure units

Depending on the connected sensor, in "MENU (Unit)" different units can be selected.
Depending on the measuring range, the selection may be limited!
EN
User unit
For units which are not covered by the units imprinted on the display, a manual setting
can be made via the user unit.
Example:
So that the unit kg/cm² can be displayed with a CPT6200 reference pressure sensor, this
must be set:
bASE user
DP user
FACt user

5.4.2 Different measuring modes

The instrument supports 3 different measuring modes for various purposes. Two
P
of these,
> 1,000 measurements/s.
5.4.2.1 Standard measurement (slow)
rate sLo
5.4.2.2 Peak value detection (Peak detection)
rate P.det
20
Downloaded from
www.Manualslib.com
kg/cm²
Torr
bar
mbar
.0000
.0000
1.0197
.7433
dEt
FASt
.
and
, operate with an increased measuring frequency of
Measuring rate 4 Hz, averaging and filter functions are active.
Application range:
Measurement of slow pressure changes and static pressures, e.g. leak
testing, atmospheric pressure measurements, etc.
Highest measuring accuracy, insensitive to disturbances, low current
consumption.
Measuring rate > 1,000 Hz; the value is displayed unfiltered.
Field of application in combination with logger function:
Measurement of pressure spikes or fluctuating pressures with a
resolution of < 1 ms. The cyclic logger function records the arithmetic
mean value, the highest and the lowest peak during the chosen time
interval.
manuals search engine
atm
at
bar
bar
.0000
.0000
.9869
1.0197
WIKA operating instructions, model CPH6300

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cph6300-s1

Inhaltsverzeichnis