5. Commissioning, operation
Menu
"MENU"
SET
Corr
SET
Corr
1) Menu cannot be selected if there is data in the logger memory. If this is to be changed, the data must first be deleted (see chapter
5.5 "Data logger").
2) This menu can only be selected if an appropriate sensor is connected to connection 1. When using a second corresponding
sensor on connection 2 then the settings are adopted.
3) Only for 2-channel version
WIKA operating instructions, model CPH6300
Downloaded from
www.Manualslib.com
Parameters
Values
▲ or ▼
sensor
daC.0
dependent
e.g. -5.00 ...
5.00 mbar
sensor
daC.I
dependent
e.g. -5.00 ...
5.00 mbar
Set Corr: Adjustment of the measurement (1-channel version)
Zero point adjustment/offset of the sensor
Offs
oFF
sensor
dependent
e.g. -5.00 ...
5.00 mbar
Slope correction for the sensor
sCAL
oFF
-2,000 ... 2,000 The measuring scale will be changed by this
Set Corr: Adjustment of the measurement (2-channel version)
Zero point adjustment/offset of sensor 1
OFS.I
oFF
sensor
dependent
e.g. -5.00 ...
5.00 mbar
Slope correction for sensor 1
SCL.I
oFF
-2,000 ... 2,000 The measuring scale will be changed by this
manuals search engine
Meaning
Zero point setting for Out = DAC: Input of the
measured value with which the analogue output
should output 0 V.
Slope setting for Out = DAC: Input of the
measured value with which the analogue output
should output 1 V.
No zero point adjustment for the sensor.
The zero point of sensor 1 will be displaced by
this value to compensate for deviations in the
probe or in the measuring instrument.
No slope correction for the sensor
factor [%] to compensate deviations of probe /
measuring instrument.
No zero point adjustment for the sensor 1.
The zero point of sensor 1 will be displaced by
this value to compensate for deviations in the
probe / measuring instrument.
No slope correction for sensor 1
factor [%] to compensate deviations of probe /
measuring instrument.
EN
1)
1) 3)
17