7 Faults / 8. Maintenance, cleaning and recalibration
Display
EN
OL
-OL
No display or
instrument is not
responding to button
press

8. Maintenance, cleaning and recalibration

Personnel: Skilled personnel
Protective equipment: Protective gloves, safety goggles
Tools: Open-ended spanner SW 27
For contact details see chapter 1 "General information" or the back page of
the operating instructions.

8.1 Maintenance

The CPG1500 precision digital pressure gauge is maintenance-free.
Repairs must only be carried out by the manufacturer.
This does not apply to the replacement of alkaline batteries.
Only use original parts (see chapter 11 "Accessories").
28
Causes
Low battery voltage, functioning is
only guaranteed for a short period
of time
Reading is significantly above or
below the measuring range = >
10 % FS
Battery is empty
Batteries inserted incorrectly
System error
Defect in the CPG1500
WIKA operating instructions precision digital pressure gauge CPG1500
Measures
Insert new alkaline batteries,
see chapter 8.1.1 "Battery
replacement".
Check: Is the pressure within the
permissible measuring range of
the sensor?
Insert new alkaline batteries,
see chapter 8.1.1 "Battery
replacement".
Ensure the correct polarity
(see chapter 8.1.1 "Battery
replacement").
Switch off the CPG1500, wait for
a short period of time, switch on
again
Send in for repair