7 Faults / 8. Maintenance, cleaning and recalibration
Fault
Instrument outside the
accuracy class
EN
Vibration of the pointer
Mechanical damage (e.g.
window, case)

8. Maintenance, cleaning and recalibration

For contact details see chapter 1 "General information" or the back page of
the operating instructions.
8.1

Maintenance

The instrument is maintenance-free.
Testing of the display function
Check the display function 1 or 2 times per year. For this, disconnect the instrument from
the process and check with a pressure testing device.
Testing of filled instruments
For filled instruments, the level must be checked once or twice every year.
The liquid level must not drop below 75 % of the instrument diameter.
Repairs must only be carried out by the manufacturer or appropriately qualified skilled
personnel.
This does not apply to the replacement of the battery or the antenna.
Only use original parts (→ see chapter 3.8 "Approved batteries and antennas").
8.2

Battery replacement

For battery replacement observe the following instructions:
Only change the batteries in a dry environment
Do not use rechargeable batteries
Only use battery model LS17500 from SAFT
Secure the battery cover with the four screws
Do not use power tools for screwing in.
Ensure the correct polarity
28
Causes
Instrument was operated
outside of permissible
performance limits.
Vibrations in the application.
Improper handling.
Impermissible loading at the
installation point (e.g. fire).
WIKA operating instructions, model PGW23.100.11
Measures
Comply with the operating
parameters.
Replace instrument.
Use instrument with case filling.
Replace instrument.