5. Commissioning, operation
5.5.1.3 Clear manual recording
If the data has already been saved, this can be cleared via the [STORE/QUIT] button:
1. Press the [STORE/QUIT] button for 2 seconds.
⇒
Accesses the "Delete" function.
EN
2. Press the [▲] or [▼] button.
⇒
Change the selection.
CLr no
CLr ALL
CLr LASt
3. Press [STORE/QUIT] button.
⇒
Confirm the selection, end of the "Delete" function.
5.5.2 Automatic recording with adjustable cycle
If the logger function,
instrument"), following the start of the logger, the measured value is automatically at the
interval of the set cycle time.
The logger cycle time is adjustable from 1 s ... 60 min.
At the moment of saving, a data set consists of the following values.
(1-channel version):
Storable data sets: 10,000
With slow measurements:
Measured value
■
CALC off
■
CALC
■
change of pressure (only for sensor connection 1)")
With fast measurements:
Arithmetic average since last save
■
CALC off
■
CALC
■
change of pressure (only for sensor connection 1)")
26
Downloaded from
www.Manualslib.com
Do not delete (cancel the process).
Delete all data sets.
Clear the last data set.
Fvnc CYCL
rAtE SLo
: Min.-Peak and Max.-Peak since last save
"active": Estimated calculation and dummy values (see chapter 5.4.5 "Rate of
rAtE FASt
: Min.-Peak and Max.-Peak since last save
"active": Estimated calculation and dummy values (see chapter 5.4.5 "Rate of
manuals search engine
, is selected (see 5.3 "Configuration of the
rAtE P.dEt
and
WIKA operating instructions, model CPH6300