Hand-Held Pressure Indicator
CPH6200-Ex (Atex-Version)
GB
16
2.8 (CLOC) Real time clock
The real time clock is needed for time-stamping each data point re-
corded within the logger function. As a result, the clock settings should
be checked regularly.
If the battery has been replaced the appropriate menu 'CLOC' will be
accessed automatically.
3. Measuring of water level - display unit [m]
In preparation.
4. Operation of logger
The device supports two different logger functions:
"Func-STOR": each time the ´Store/Quit´ key is pressed a data point
will be recorded.
"Func-CYCL": measurements will automatically be recorded at a
defined interval, set within the logger menu 'CYCL'
until the logger is stopped or the logger memory
becomes full. Logging is started by pressing the
´Store/Quit´ key for 2 seconds.
The logger records up to 3 measurement values for each interval:
Instantaneours or mean value (depending on logger setting, see
below), Min-Peak and Max-Peak.
To evaluate the data, WIKA's Data logger evaluation software GSOFT
(V 1.7 or higher) must be used. The software also allows easy configu-
ration and operation of the logger.
If the logger is selected (Func-STOR or Func-CYCL) the hold function
is deactivated; the ´Store/Quit´ key is then used solely for the opera-
tion of the logger functions.
Min- and Max-Peak are, respectively, the minimum and the maximum
measured values during the last measurement interval. Therefore both
the instantaneous pressure values and the measured fluctuations can
be better analysed.
WIKA Operating Instructions Hand-Held Pessure Indicator