Herunterladen Inhalt Inhalt Diese Seite drucken

Ottobock Genium X3 3B5-3 Gebrauchsanweisung Seite 59

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
CAUTION
Failure to observe activated mute mode (silent mode)
Falling due to unexpected product behaviour because of changed damping behaviour.
The following feedback signals are deactivated when mute mode is activated:
>
Long vibration signal if the hydraulic unit overheats.
>
Beep and vibration signal to confirm that the movement pattern has been recognised
(switching to a MyMode/basic mode with movement pattern).
>
Beep and vibration signal to indicate successful switching to a MyMode/basic mode.
>
Beep and vibration signal upon successfully switching to deep sleep mode.
Before activating mute mode, note that these feedback signals will be deactivated. For more
information about mute mode, see the section "Mute mode" (see Page 81).
After switching to a MyMode/basic mode, verify the changed damping settings.
Ensure that you stand securely during all switching processes.
To deactivate mute mode, connect and then disconnect the battery charger.
CAUTION
Independent user manipulation of system components
Falling due to breakage of load-bearing components or malfunction of the product.
Manipulations to the product other than the tasks described in these instructions for use are
not permitted.
The battery may only be handled by Ottobock authorised, qualified personnel (no replace­
ment by the user).
The product and any damaged components may only be opened and repaired by authorised,
qualified Ottobock personnel.
CAUTION
Mechanical stress on the product
>
Falling due to unexpected product behaviour as the result of a malfunction.
>
Falling due to breakage of load-bearing components.
>
Skin irritation due to defects on the hydraulic unit with leakage of liquid.
Do not subject the product to mechanical vibrations or impacts.
Check the product for visible damage before each use.
CAUTION
Use of the product when battery charge level is too low
Falling due to unexpected behaviour of the prosthesis because of changed damping behaviour.
Check the current charge level before use and charge the prosthesis if required.
Note that the operating time of the product may be reduced at low ambient temperatures or
due to ageing of the battery.
CAUTION
Risk of pinching in the joint flexion area
Injuries due to pinching of body parts.
Ensure that fingers/body parts or soft tissue of the residual limb are not in this area when
bending the joint.
59

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Genium x3 3b5-3-st

Inhaltsverzeichnis