- Connect the handpiece cable to the connector
(Fig. 1-7).
- Check that the device is sufÀ ciently close to the
working area to avoid having to stretch the cables.
If necessary, move the device closer.
- Hang the bottle or pouch containing physiological
saline solution from the bracket.

4.4 FIRST START-UP

Before using the device for the À rst time, it is vital
that you service and/or sterilize all equipment in
accordance with the procedures deÀ ned in section 11.
V - IRRIGATING SOLUTIONS
The device is not designed to administer medical subs-
tances and shall only be used with bottles or pouches
containing no more than 1 litre of physiological saline
solution or sterile water.
VI - IRRIGATION LINES
The device shall be used with SATELEC irrigation lines.
VII - SETTINGS / PARAMETERS
/ INTERFACE

7.1 START-UP

Important:
It is recommended that you wait 4 seconds between
stopping and restarting the device.
At start-up, the device performs an auto-check
sequence and the screen displays programme d1
details.

7.2 PARAMETERS

The parameters are stored during the conÀ guration
of each programme and are recalled each time you
select a programme.

7.3 FLUSH / PRIME

The Á ush/prime function is controlled using the
keyboard.
Press button
(Fig. 3-3) to activate the Á ush/
prime function.
This function remains active while the button is
pressed down.
Flush can be activated during use of the device. The
Á ow rate during Á ush is 120ml/min.

7.4 IRRIGATION

a) Setting the irrigation Á ow rate
Set the irrigation Á ow rate by pressing the buttons
and
(Fig. 3-1 and 2).
Each press adjusts the Á ow rate by + or – 10ml/min.
The value set is then displayed on the LCD screen (Fig.
3-7) in addition to a bar graph.
The Á ow rate is stored in the programme being used.
The Á ow rate can be altered during use of the device.
b) Deactivating irrigation
To deactivate irrigation, press the
3-2) until 0ml/min is displayed. The irrigation bar
graph (Fig. 3, item 7) is then blank.

7.5 ULTRASOUND FUNCTION

Important:
The irrigation parameters can be customised by the
user for each programme.
The ULTRASOUND function can be adjusted when
the handpiece is not connected to the device.
Programme d1 is the most powerful setting.
Select the programme to be used by pressing buttons
,
,
or
the symbol
on the pedal (Fig. 4-2)
The selected programme appears on the LCD screen
as d1, d2, d3 or d4 in addition to an associated bar
graph (Fig. 3-8)
button (Fig.
(Fig. 3-5) or by pressing
9