Immediate display of centrifugation data after switch-on.


Switch on the power supply. (Switch setting ).

The first time the display changes optically, (inverse display) press any key and keep it held down. The
centrifugation data is displayed immediately.
34

Cooling (only in centrifuges with cooling)

The temperature set-point can be adjusted from -20°C to +40°C / -4°F to +104°F. The lowest obtainable temperature
is dependent on the rotor (see Chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories").
34.1

Standby-cooling

When the rotor is at a standstill, with the lid and hatch closed, the centrifuging chamber is cooled to the preselected
temperature if this is lower than 20°C / 68°F.
During the standby cooling, the preset temperature is displayed.
34.2

Precooling of the rotor

For quick precooling of the unloaded rotor and the accessories, it is a good idea to do a centrifugation run with
the continuous operation setting and a speed of about 20% of the maximum rotor speed.
A centrifugation run for pre-cooling the rotor can only be started via the interface (key position "LOCK 2"). The
functions of the various key positions are described in the "Key-operated switch" chapter.

Start the centrifugation run via the interface. The LED in the
it is subsequently lit.

After the time elapses or if the centrifugation run is aborted via the interface or by pressing the
the run-down occurs with the activated run-down parameters. The run-down parameter is displayed e.g.
The right-hand LED in the
key and the right LED in the
During the centrifugation run, the rotor speed or the RCF value, the temperature in the centrifuging chamber and the
remaining or elapsed time are displayed.
34.3

Time-controlled cooling

If necessary it can be set so that cooling begins after a certain amount of time once centrifugation run has been
started.
The delay time can be set in 1-second increments from 15 to 900 seconds. If no time delay is required, set to "0". No
delay time is set ex works.
When the rotor is at a standstill, the delay time can be set as follows:
This setting can only be made via the operating unit (key position "0" or "LOCK 2"). The functions of the
various key positions are described in the "Key-operated switch" chapter.
You can scroll backwards through the menu by pressing the
The operation can be cancelled at any time by pressing the
stored.

Keep the
key held down for 8 seconds.
PROG
After 8 seconds,  Machine Menu  appears in the display.

Press the
key as often as necessary until the ->Settings is displayed.
PROG

Press the
key. The SOUND / BELL = off/on is displayed.
START

Press the
key as often as is needed until the delay time is displayed, e.g. Cool acc time = 0.
PROG

Use the adjusting knob

0 = no delay time

Press the
key to save the setting.
START
Briefly, Store Settings ... is displayed as confirmation, followed by -> Settings.

Press the
key once to exit the "Settings" menu or press the
OPEN / STOP
"Machine Menu".
74/83
key is lit. Once the rotor has come to a standstill, the LED in the
OPEN / STOP
key switch off and the left LED in the
OPEN / STOP
to set the value you want.
key blinks until the rotor has been imported,
START
OPEN / STOP
key.
T/°C
key. In this case, the settings are not
OPEN / STOP
key twice to exit the
OPEN / STOP
key,
OPEN / STOP
9.
START
key lights up.