WaveWriter Alpha™and WaveWriter Alpha™ Prime Implantable Pulse Generator
Connecting the Lead, Lead Extension, Splitter, or Connector to the 16
Contact IPG
Note: Lead Extensions, Adapters, and Splitters are not full body MRI scan eligible.
IPG ports are labeled as follows (see Figure 3).
C
Figure 3.
IPG Ports C and D
• Linear Leads connect to IPG ports C or D.
• For the Infinion 16 Lead, connect the Splitter tail with laser-etched bands (Contacts 1-8 of the
Infinion 16 Lead) to the left Port C and the unmarked Splitter tail (Contacts 9-16 of the Infinion 16
Lead) to the right Port D.
• For the Infinion CX Lead or Infinion Pro Lead, connect the tail with the single marker band
(Contacts 1-8 of the Infinion CX Lead or Infinion Pro Lead) to the Port C and the tail with two
marker bands (Contacts 9-16 of the Infinion CX Lead or Infinion Pro Lead) to the right Port D.
• For the Artisan and Artisan MRI 2x8 Surgical Leads, connect the left side to the left Port C. Connect
the right side (the laser-etched tail), Contacts 9-16, to the right port D.
Implantable Pulse Generator
92959607-02