Sensitivity
The sensitivity of the force gauge can be set here in mV/V. Once you have selected this item, a
popup will appear. Here, you can select the digit to modify with the ◄ and ► keys and with the
▲ and ▼ keys, you can increase and reduce the value of the digit. Once the value is set, it can
be saved by pressing the OK key. Press the back key to return to the Sensitivity menu or discard
the set value as long as the popup is visible.
Attention:
Do not change this parameter when using the PCE-DFG X with internal load cell.
5.2.5.3.1.4

Zero offset

The zero offset of the force gauge can be set here. When the item is selected, a popup will appear
and continuously display the reading of the force gauge in mV. In order to set the offset, the force
gauge has to have no weight or force applied and once the value displayed has stabilized, press
the OK key to store that value as zero offset.
To return to the calibration menu, press the back key.