5.2.5.3.2
Gravity value
This parameter is used to adjust the measured values by the gravity value the user is experiencing
and the gravity value used for the applied calibration. Once you have selected this item, a popup
will appear. Here, you can select the digit to be modified with the ◄ and ► keys and with the ▲
and ▼ keys, you can increase and reduce the value of the digit. Once the value is set, it can be
saved by pressing the OK key. Press the back key to return to the calibration menu or discard the
set value as long as the popup is visible.
5.2.5.3.3
High resolution
This item allows you to select whether another decimal is to be displayed or not. This extra
decimal will be more unstable than the others. Once this item is selected, you can use ▲ and ▼
to set it to On or Off. Once the item is set, press OK to save the selection.
© PCE Instruments
37