Further functions

7.1

Auto zero

If you start a measurement and initiate rotation without immersing the rotor into a fluid and the display
does not show zero, you have to use the auto zero function.
To do so, follow these steps:
Turn on the device by holding the Enter button for 2 seconds.
When the information screen appears, press the Down button and then the Enter button shortly, one after
the other.
Now you see this screen:
Press Enter to get to the next screen:
Now, remove the spindle/rotor and press Enter again to start the zeroing procedure. During the zeroing,
the display shows the following screen:
After the zeroing was successful, the viscometer automatically goes back to the information screen.
7.2

Change the measuring unit

To select the measuring unit (you can choose between mPa s and dPa s), turn on the viscometer by
pressing and holding the Enter button for 6 to 10 seconds. Once the information screen appears, press
the Up button followed by the Down button. Now you get to the following screen:
Press Enter to open the unit selection screen:
Here you can select the desired measuring unit (mPa s or dPa s) by using the Up and Down buttons.
After that, press Enter to confirm. You will be redirected to the information screen automatically.