5.2.5.3.1.1
Maximum load unit
The maximum load unit determines the units for the maximum load value and the calibration
weights of the calibration curves. The available options are kg, N and Nm. If a torque unit is
selected, the units regarding weights or pressure will not be displayed in the Units submenu in
the Measurement menu, and vice versa. Press OK to start selecting the unit and use ◄ and ►
to choose between the options. Confirm the option with the OK key and use the back key to stop
changing the option and restore the original.
5.2.5.3.1.2
Maximum load
Depending on the maximum load unit selected, the maximum load value has to be entered in that
unit. Once you have selected this item, a popup will appear. Here, you can select the digit to
modify with the ◄ and ► keys and with the ▲ and ▼ keys, you can increase and reduce the
value of the digit. Once the value is set, it can be saved by pressing the OK key. Press the back
key to return to the sensitivity menu or discard the set value as long as the popup is visible.
© PCE Instruments
35