ECOtwin LCD / LED
Battery Disposal and Handling
The capacity of internal battery is gradually decreased over time and usage. Therefore the operation
time by the battery can be decreased. If the operation time is not long enough, please contact service
center and change the battery. If this system is used with not sufficient operating time by the internal
battery, it is possible to be shut down the system because of the lack of the internal battery's
capacity. This situation can cause not intended stop of measuring and monitoring function.
CAUTION
When disposing of internal Li-ion battery, adhere to all applicable laws regarding
recycling. Avoid storing battery above 140°F. If clothing or skin comes in contact with
material from inside the battery, immediately wash with plenty of clean water.
The internal battery must be handled by the company's technician only. Do not attempt to
open the ECOTWIN.
Maintenance
The ECOTWIN monitor and accessories require no periodic calibration or adjustment. The
recommended interval for performing hipot and leakage testing is once per year.
Disposal of the ECOTWIN
When disposing of the ECOTWIN, adhere to all applicable laws regarding recycling. If you are not
able to dispose the ECOTWIN or you need a help for disposing the ECOTWIN, please contact us.
In the case of there are no appropriate ways to dispose, we will pick up the ECOTWIN for you.
User manual
62