The basic operation of control knob for configuration settings is as follows.
Activity
Press
Rotate
Press
Rotate
Press
Setting Alarm Upper Limit / Lower Limit
Activity
Knob Rotate
Knob Press
Knob Rotate
Knob Press
Setting Dop2 Offset
When ultrasound trace separation is enabled, the trend data for ultrasound channel 2 is shifted up by
[+10], [+20], [+30] or [+40]
rate trends when both heart rates are similar. The heart rate value shown in the numeric frame is not
affected.
Activity
Knob Rotate
Knob Press
Knob Rotate
Knob Press
Setting FM (Fetal Movement) Graph
Activity
Knob Rotate
Knob Press
Knob Rotate
Knob Press
Setting Print Speed
Activity
Knob Rotate
User manual
Desired Result
Enter the configuration setting mode.
Select the setting item
Select the parameter to change.
Change the set value
Store the new value.
Desired Result
Select "UPPER LIMIT" or "LOWER LIMIT"
Select this parameter for change.
Change the desired value.
The list below shows the values that are available for each
parameter:
Heart Rate Upper Limit { (Heart Rate Lower Limit +10) ~ 240 BPM, 5
BPM step}
Heart Rate Lower Limit {30 ~ (Heart Rate Upper Limit-10) BPM, 5 BPM
step}
Store the new value.
BPM in printing. This feature is provided to clearly see separate heart
Desired Result
Select "DOP2 OFFSET"
Select this parameter for change.
Change the desired value.
Store the new value.
Desired Result
Select "FM GRAPH"
Select this parameter for change.
Change the desired value [ON/OFF].
Store the new value.
Desired Result
Select "SPEED"
23