Herunterladen Inhalt Inhalt Diese Seite drucken

Belts; Contacting Components And Characteristics; Description Of Cidex Tm - Medical Econet Ecotwin Lcd Bedienungsanleitung

Zwilings-farbmonitor
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

ECOtwin LCD / LED

Belts

Wash soiled belts with soap and water.
CAUTION
The water temperature must not exceed 60°C (140°F).

Contacting components and characteristics

Contacting
component
DOP & UC
Housing
Strain gauge
sensor housing

Description of Cidex TM

TM
1. Cidex
is FDA-cleared for use in the United States. Therefore we suggest that the disinfection
effect using Cidex
2. FDA-Cleared Sterilants and High Level Disinfectants with General Claims for Processing
Reusable Medical and Dental Devices – March 2009
(www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/ReprocessingofSingle-
UseDevices/UCM133514)
Manufacturer
K924434 Cidex™ Activated Dialdehyde Solution
Johnson &
Johnson
Medical
Products
User manual
Material
ABS AF-302
Estane S385A
TPU
TM
is valid.
Active Ingredient
2.4%
glutaraldehyde
Usage
Must be cleaned and disinfected
Reusable
prior to use
Must be cleaned and disinfected
Reusable
prior to use
Sterilant
High Level Disinfectant Contact
Contact
Conditions
Conditions
45 min at 25°C
10 hrs at 25°C
14 days Maximum Reuse Contact
14 days
conditions based on literature
Maximum
references.
Reuse Contact
conditions based
on AOAC
Sporicidal
Activity Test
only.
Disinfection
56
Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Ecotwin led

Inhaltsverzeichnis