ECOtwin LCD / LED
Detail Procedure
Explain procedure to the patient.
①
Place a probe belt under the patient
②
Turn the monitor power on. The power switch is located on the rear panel. The green indicator
③
located under the left side of the printer door illuminates when the power on.
Connect the transducer plug to "UC" connector located on the underside of the front cover.
④
Note: When connector or re-connecting the UC probe to the monitor's UC connector,
you must wait at least 10 seconds before depressing the UC reference [
Briefly depress the UC reference [
⑤
Position UC probe on the maternal abdomen over the uterine fundus or where there is the least
⑥
maternal tissue and the contractions are strongly palpated.
Connect each end of the belt to the transducer by inserting the transducer button through a
⑦
buttonhole on the strap. Select a buttonhole that ensure a comfortable fit and holds the transducer
securely in the place.
Between contractions, depress the UC reference [
⑧
The monitor is now ready to begin monitoring.
⑨
If not already activated, depress the [
The recorder plots the UC on the paper strip chart.
CAUTION
The probe belt may cause allergy or skin side effects to patient, if it is used so long time.
User manual
] pushbutton to set the UC baseline at 10.
] button again. This set UC baseline to 10.
] pushbutton located on the front panel of the monitor.
] button.
53