CCV (Cross-channel Verification) Function

CCV Function Setting

During Self-Test, press Control Knob 3 times then, there will be pop up message "CCV Select".
Rotate the control knob to Left/Right way then, it is available to change On/Off mode for CCV
Function.
Once CCV function is set, the function will be maintained when restart the equipment.
After setting CCV Function, please know button 3 times then, the equipment will be restarted.
When 1 st fetal heart rate is the same with 2 nd fetal heart rate (CCV) during using 2ea Doppler to the
patient,
icon will be appeared and alarm will be activated. (If one probe is removed or it is not
CCV, alarm will be stopped and the equipment will work normally.)