ECOtwin LCD / LED
ECOTWIN Monitor Control Knob and system setting
The Control Knob is the primary method of adjusting parameters and navigating through the menu
system. If the knob is rotated to the CW (clockwise) or CCW (counterclockwise) while in a menu,
the cursor moves throughout the items within the menu. This process is used to select a menu item
for modification. The knob is then pressed to select this item for editing.
Once a menu item has been selected for editing, the knob is rotated to scan through the available
choices for this parameter. Pressing the knob stores the new value temporarily.
Pressing the knob when "ESC" is selected will exit the present menu and save the changed value.
The screen will back on monitoring mode when ESC is pressed on SETUP mode and relevant menu
screen, and the screen will back to SETUP mode in case ESC is pressed on HELP Screen.
The monitor has several configuration settings that the user can change. Some of these settings are
reset to the default value each time the monitor is powered down. Other parameter settings are saved
in the monitor until the next time they are changed. These parameters are unaffected when the
monitor is powered down. A complete list of these parameters is shown below.
Fetal Heart Rate Upper Alarm Limit
Fetal Heart Rate Lower Alarm Limit
User manual
Fig. 4.4 System setup menu
Configuration parameter
Dop2 Trace Separation
(Dop2 Offset)
FM Graph
Record Paper Speed
Auto Printing
Patient Name
Patient ID
Date
Auto Save
Time
Language
Paper
Factory Default
190 BPM
110 BPM
0 BPM
OFF
3 cm/min
0 MIN
blank
Date/Sequential number
YY/MM/DD
OFF
HH:MM:SS
English
FS151-90-80R-01