Specifications
Physical Characteristics
Dimensions
Weight
Safety
Standard
Classification
Mode of Operation
Protection against electric
shock
Protection against ingress
of water
Power
External
Internal
Power Dissipation
Environment
Operating temperature
Operating humidity
Storage temperature
Storage humidity
Altitude
Pressure
Doppler ultrasound FHR monitoring
BPM Range
Accuracy
Leakage
Isolation
Uterine Contraction (TOCO) monitoring
UC range
Resolution
Accuracy
Leakage
Isolation
User manual
9.6cm(H) x 32.6 cm(W) x 27.6cm(D)
Approx. 5.5 kg
EN 60601-1, EN60601-2,
Class I, Internal Powered Equipment
Continuous operation
Type BF applied part
IPX8 (Dop/UC probe)
Power adapter
Battery
AC-powered
Battery
powered
10°C ~ 45°C (50°F ~ 110°F)
5 ~ 85%, Non condensing
-20°C ~ 60°C (-4°F ~ 140°F)
0 ~ 95%, non-condensing
0 ~ 2 000 m(0 ~ 6 561.68 ft)
79.051 kPa ~ 101.325 kPa
30 ~ 240 BPM
± 2% of range
<10 µA @ 264 VAC applied to transducer
>4 kV RMS, Type BF applied part
0-99 relative units
1 Count
± 1% relative unit
<10 µA @ 264 VAC applied to transducer
>4 kV RMS, Type BF applied part
EN 60601-2-37
Input: AC 100 ~ 240 V, 50/60 Hz
Output: DC 18V, 2.8A
14.8V, Li-ion
80 VA, maximum
80 VA, maximum