1.5.1 Calibration
– For proper operation, never adjust the span before
completing zero adjustment. Performing these operations
out of order will cause the calibration to be faulty.
– If the intended operation is at high altitudes, the reading
will be lower than the reading at sea level (reduced partial
pressure). A new span calibration is recommended if the
altitude or the ambient pressure is changed. The factory
calibration is set to sea level.
– Dräger recommends calibrating instruments with target
gas. This method is more accurate than a surrogate gas
calibration. A surrogate gas calibration may only be
performed as an alternative if a target gas calibration is
not possible.
– Methane and hydrogen should be calibrated with target
gas only and not calibrated with a surrogate gas.
Only for Polytron 5310 IR:
– Only for DSIR (the user must also read the Instructions for
Use for DrägerSensor IR, part number 9023981)
1.6
During operation
– After modifying parameters with the PolySoft PC software
or another software, verify all parameters by downloading
from the instrument or by checking them on the Polytron
5xx0.
– IP ratings do not imply that the equipment detects gas
during and after exposure to those conditions. In case of
dust deposits and submersion/jet water, check the
calibration and the proper functioning of the device.
Risk of ignition of flammable or explosive atmospheres!
– CAUTION: High off scale readings may indicate an
explosive concentration.
(Only for instruments measuring combustible
atmospheres.)
– WARNING: Do not open when energized
Polytron 5200 CAT-specific
– When exposed to a directed flow of air mixed with gas, the
measured values can be increased. See the
corresponding sensor data sheet for deviation values.
1.6.1 Maintenance
– The maintenance intervals must be established for each
individual installation. Depending on safety considerations
and application specific conditions the instrument is used
in, these might need to be shortened.
– Refer to the maintenance section in the transmitter
instructions for use.
2
Conventions in this document
2.1
Meaning of the warning notes
The following alert icons are used in this document to provide
and highlight areas of the associated text that require a
greater awareness by the user. A definition of the meaning of
each icon is as follows:
8
Alert icon Signal word
WARNING
CAUTION
NOTICE
3
Description
3.1
Product overview
Refer to the figures on the fold-out page
A 1
Enclosure lid
A 2
Set screw (2 mm Allen screw)
A 3
Relay board
A 4
PCB unit
A 5
Enclosure bottom
A 6
Port for Sensor
A 7
Feed-through cable
A 8
Docking station
A 9
Explosion proof instrument
A 10 Explosion proof instrument with increased safety wir-
ing compartment (Docking Station)
A 11 Explosion proof instrument to be extended with mount-
ing spacer for Polytron 57x0 IR
A 12 Mounting spacer (only Polytron 57x0 IR)
B
Calibration set-up
C
Polytron 5100 EC
D 1
Polytron 5200 CAT
D 2
Polytron 5310 IR
E 1
Polytron 57x0 IR with splash guard
E 2
Polytron 57x0 IR without splash guard
3.2
Feature description
Polytron 5xx0 instruments are housed in a rugged stainless
steel or aluminum enclosure for indoor and outdoor
applications.
Instructions for use
Consequences in case of
nonobservance
Indicates a potentially hazard-
ous situation which, if not
avoided, could result in death
or serious injury.
Indicates a potentially hazard-
ous situation which, if not
avoided, could result in injury. It
may also be used to alert
against unsafe practices.
Indicates a potentially hazard-
ous situation which, if not
avoided, could result in dam-
age to the product or environ-
ment.
|
Dräger Polytron 5000 Series

Werbung

loading