I LIMB HAND COVERS
Donning a Cover
To don a cover, bring i-Limb hand into the position shown in Figure 16 and turn the hand off. The position can also
be achieved using the don/doff quick grip.
Donning the i-Limb Skin Active, i-Limb Skin Contour and i-Limb Skin Natural covers:
1. Line up cover with the fingers of the i-Limb hand, and slide cover downwards.
2. When the fingers are mostly donned, pull the thumb opening over the thumb (Fig. 17).
3. Slide the remainder of the cover over the i-Limb hand (Fig. 18).
4. Make sure each digit tip is fitted to the cover.
5. Do not pull the cover too tightly over the hand.
Donning the i-Limb Skin Match cover:
1. Spray the outer surface with isopropyl alcohol (Fig. 19).
2. Invert the cover to the level of the finger openings (Fig. 20).
3. Ensure the fingers are straight and not bunched.
4. Align the cover finger holes with the digits of the hand (Fig. 21).
5. Pull down onto digits of the i-Limb hand.
6. Place thumb opening over thumb digit.
7. Pull the cover carefully over the remainder of the i-Limb hand, without applying too much pressure on the
thumb (Fig. 22).
8. Check the cover for bunching and make sure the cover tips are fully fitted to the digit tips.
9. Check the function of the i-Limb hand and ensure full opening and closing is possible and the digit tips align.
Doff ing all Covers
1. Bring the i-Limb hand into the same position as for donning and switch off/ to standby.
2. Pull the cover upwards on each digit to release.
3. Carefully ease the full cover off, without applying too much pressure on the thumb.
4. Continue to pull cover upwards, until fully removed.
CLEANING
Cleaning instructions
The i-Limb hand covering, electrode surface and magnetic charger port can be cleaned with a soft damp cloth and
mild soap.
Clean the electrode surface regularly.
Clean the cover with isopropyl alcohol once a week to help with disinfection.
NOTE: do not submerge the i-Limb hand, cover, electrode surface or magnetic charger port in water. Do not use
any strong chemicals.