• Do not use the siderails as a push or pull device. Always move the product using the integrated handles in the
headboard and footboard.
• Always remove the patient helper before you transport the product.
• Do not use the patient helper as a push or pull device.
• Do not use the oxygen bottle holder as a push or pull device.
• Do not use the IV pole as a push or pull device.
• Always make sure that the IV pole is at a low height during transport.
• Do not transport the product on inclines greater than 6 degrees (10% grade).
• Always clear the product of obstacles before you use motion functions.
• Always confirm that all persons and equipment are away from the area below and around the Fowler before you activate
the CPR release handle. The CPR release handle is for emergency use only.
• Do not load the IV pole above the safe working load of 17.6 lb (8 kg).
• Do not load an individual IV pole hook above the safe working load of 8.8 lb (4 kg).
• Always secure the lifting pole in the mounting bracket before you adjust the patient helper.
• Always check that the patient helper mounting bracket is secure before use.
• Do not load the standard oxygen bottle holder (300900450050) above the safe working load of 15 lb (6.8 kg).
• Do not load the large oxygen bottle holder (300900450150) above the safe working load of 40 lb (18.1 kg).
• Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and motors.
Remove the occupant from the product, clean up the fluid, and inspect the product. Fluids can cause unpredictable
operation and decreased functionality of any electrical product. Do not return the product to service until dry and tested
for safe operation.
• Always wipe down with clean water (or 70% isopropyl alcohol, if using V V i i r r e e x x ® TB) and dry each product after
disinfecting. Some disinfectants are corrosive in nature and may cause damage to the product. If you do not rinse and
dry the product, you may leave a corrosive residue on the surface of the product. This corrosive residue could cause
premature degradation of critical components. Failure to follow these disinfecting instructions may void your warranty.