Conducted RF
IEC 61000- 4-6
Radiated RF
IEC 61000-4-3
N N o o t t e e - - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
N N o o t t e e - - The ISM (Industrial, Scientific, and Medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
a
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the P P r r o o C C u u i i t t y y bed series is used exceeds the applicable
RF compliance level above, the P P r r o o C C u u i i t t y y bed series should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the P P r r o o C C u u i i t t y y bed
series.
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 Vrms.
b
3009-009-005 Rev AD.2
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
Portable and mobile RF
communications equipment
should follow the guidance
in the table titled
"Recommended separation
distances between portable
and mobile RF
communication equipment
and the P P r r o o C C u u i i t t y y bed
series." If the mobile service
is not listed in the table, the
recommended separation
distance should be
calculated from the equation
appropriate for the
frequency of the transmitter.
Recommended separation
distance
D=(2) (√ P )
3 Vrms
where P is the maximum
3 V/m
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
survey
the compliance level in each
frequency range
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
, should be less than
a
.
b
75
EN