position, Bed Exit zone, sensitivity, i i B B e e d d W W a a t t c c h h enablement, bed motion lock, and bed scale observation. The desired bed
parameters will be set by clinicians at bedside. i i B B e e d d W W i i r r e e l l e e s s s s with i i B B e e d d W W a a t t c c h h is intended for use only with specific
enabled Stryker beds that are verified and validated with the i i B B e e d d W W i i r r e e l l e e s s s s software and is not intended to provide bed
status information for non-Stryker beds. Patient health information is not communicated or stored.
The P P r r o o C C u u i i t t y y bed series has not been evaluated for compliance to bed standard BS EN 50637. This product is not
intended for use with pediatric patients or adult patients with atypical anatomy in markets that recognize this bed standard
for market authorization.
This product is not intended for:
Behavioral health patient use
Oxygen rich environments
Sterile environments
Home care or non-institutional long-term care facility settings
C C l l i i n n i i c c a a l l b b e e n n e e f f i i t t s s
Patient treatment, patient positioning, and diagnostic
C C o o n n t t r r a a i i n n d d i i c c a a t t i i o o n n s s
None known.
E E x x p p e e c c t t e e d d s s e e r r v v i i c c e e l l i i f f e e
The P P r r o o C C u u i i t t y y bed series has a 10 year expected service life under normal use conditions and with appropriate periodic
maintenance.
The backup batteries have a two year expected service life under normal use conditions.
D D i i s s p p o o s s a a l l / / r r e e c c y y c c l l e e
Always follow the current local recommendations and/or regulations governing environmental protection and the risks
associated with recycling or disposing of the equipment at the end of its useful life.
S S p p e e c c i i f f i i c c a a t t i i o o n n s s
W W A A R R N N I I N N G G - - Always use Stryker approved support surfaces that have been tested for compatibility with the product frame
to avoid the risk of patient entrapment.
Safe working load
N N o o t t e e : : Safe working load indicates the sum of the
occupant, accessories, and mattress weight.
Maximum patient weight
Mass of equipment with safe
working load
Product weight
3009-009-005 Rev AD.2
Standard
Z Z o o o o m m (ZM option)
Standard
Z Z o o o o m m (ZM option)
550 lb
500 lb
1125 lb
1235 lb
575 lb
685 lb
249.5 kg
226.8 kg
510.3 kg
560.2 kg
260.8 kg
310.7 kg