f) ME EQUIPMENT and ME SYSTEMS with a d.c. power input intended for use with a.c.-to-d.c. converters shall be tested
IMMUNITY TEST LEVELS are applied to the a.c. power input of the converter.
g) Applicable only to the device connected to single-phase a.c. mains.
h)
i) ME EQUIPMENT and ME SYSTEMS with RATED input current greater than 16 A / phase shall be interrupted once for
250/300 cycles at any angle and at all phases at the same time (if applicable). The device with battery backup shall
resume line power operation after the test. For ME EQUIPMENT and ME SYSTEMS with RATED input current not
j) ME EQUIPMENT and ME SYSTEMS that does not have a surge protection device in the primary power circuit may be
tested only at ± 2 kV line(s) to earth and ± 1 kV line(s) to line(s).
k) Not applicable to CLASS II the device .
l) Direct coupling shall be used.
m)
n)
o) Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal to 16 A / phase and ME
EQUIPMENT and ME SYSTEMS with RATED input current greater than 16 A / phase.
p) Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal to 16 A / phase.
q)
r)
voltage within the range.