11.1.3
11.1.3.1
The customer or the user of the device should ensure that it is used in such environment.
Electrical fast
transients /
IEC 61000-4-4
bursts
a) l) o)
Surges
a) b) j) o)
IEC 61000-4-5
Line-to-line
Surges
a) b) j) k) o)
Line-to-
IEC 61000-4-5
ground
Conducted
disturbances
IEC 61000-4-6
induced by RF
c) d) o)
Voltage dips
IEC 61000-4-11
f) p) r)
Voltage
interruptions
IEC 61000-4-11
f) i) o) r)
a) The test may be performed at any one power input voltage within the device RATED voltage range. If the device is
b) All the device cables are attached during the test.
c)
d)
The quality of the network
±2 kV on contact
voltage should be that of a
typical commercial or hospital
frequency
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Portable and mobile RF
communication devices
3 V
m)
should not be used near any
6 V
in the ISM bands
m)
respect the recommended
n)
e)
equation applicable at the
frequency of the transmitter.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
h)
The quality of the network
voltage should be that of a
h)
typical commercial or hospital
environment.
29