Herunterladen Inhalt Inhalt Diese Seite drucken

B. Braun Aesculap Spine activ C FW857R Gebrauchsanweisung Seite 7

Einsetzinstrument
Vorschau ausblenden Andere Handbücher für Aesculap Spine activ C FW857R:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 18
Assembling
Note
Account for the size of the implant when selecting the
insertion instrument!
Remove irrigation connector 5 from the handle
end of the insertion instrument 1.
Ensure that the clamping sleeve 7 on the insertion
instrument 1 has been loosened.
Turn clamping sleeve 7 counterclockwise, until it is
no longer in connection with the connecting
piece 6 of the insertion instrument 1.
Connect the appropriate spacer for the intra-oper-
atively selected implant height 2 with the insertion
instrument 1.
Insert the spacer 2 in the hole of the connecting
piece. Observe the "CRANIAL" and "CAUDAL" mark-
ings on the spacer 2 and insertion instrument 1. At
the same time, keep button 9 "push to clean" on
the handle of the insertion instrument 1 pressed
down.
Insert the spacer 2 into the insertion instrument as
far as it will go 1.
Release button 9 "push to clean". By pulling the
spacer axially 2 check that it is locked in the inser-
tion instrument 1 correctly.
By turning the adjusting wheel 8 counterclock-
wise, adjust the depth stop to the minimum inser-
tion depth.
1
8
7
6
Fig. 1
9
2
5

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis