activ C-Einsetzinstrument
Inserting the implant
Damage to vertebral body end
plates when the intervertebral
disk prosthesis is tapped in!
WARNING
Carefully, under X-ray control, insert the disk pros-
thesis into the intervertebral disk space. The supe-
rior prosthesis plate (plate with the spikes) must be
oriented cranially, the inferior prosthesis plate (the
plate with the fin) caudally.
If necessary impel the implant with light hammer
taps on the proximal end of insertion instrument 1.
Compression of the spinal
canal and of other posterior
elements caused by the inter-
WARNING
vertebral disk prosthesis
inserted too deeply!
If necessary, adjust the position of the depth stop
with the adjusting wheel 8, until the desired posi-
tion is reached.
If reduction is necessary, screw the repositioning
instrument 4 on the proximal end of the insertion
instrument 1 and carefully establish the desired
position of the implant using a slotted hammer.
4
Apply proper care when
tapping in the interverte-
bral disk prosthesis.
Prior to inserting the inter-
vertebral disk prosthesis,
adjust the depth stop to
minimum insertion depth.
Introduce the disk prosthe-
sis into the intervertebral
space under X-ray control.
Decoupling the implant
Note
To avoid damage to the implant, do not couple an inser-
tion instrument that already has been decoupled once.
Loosen clamping sleeve 7 on the insertion
instrument 1 by turning counterclockwise and
removing the insertion instrument 1.
If the clamping sleeve 7 is difficult to loosen, insert
key 3 in one of the lateral holes on the clamping
sleeve 7 to loosen the clamping sleeve 7.
On the intraoperative X-ray image, in AP and lat-
eral view, check the size, height and position of the
intervertebral disk implant.
Disassembling
Decouple the spacer 2 by pressing the button 9
"push to clean" on the handle of the insertion
instrument 1 and at the same time pulling it out
axially towards the working end of the insertion
instrument 1.
If necessary, unscrew repositioning instrument 4
from the insertion instrument by turning counter-
clockwise 1.
Before reprocessing, ensure that the clamping
sleeve 7 is sitting loosely on the instrument shaft.