Mobilize non-rigid components, such as set screws,
links, etc. during cleaning.
►
Thoroughly rinse through these components with
the cleaning disinfectant solution (at least five
times), using a disposable syringe.
Phase II
►
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
►
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
►
Drain any remaining water fully.
Phase III
►
Fully immerse the product in the disinfectant solu-
tion.
►
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
►
Rinse lumens at least 5 times at the beginning of
the exposure time using an appropriate disposable
syringe. Ensure that all accessible surfaces are
moistened.
Phase IV
►
Rinse/flush the product thoroughly (all accessible
surfaces).
►
Mobilize non-rigid components, such as set screws,
joints, etc. during final rinse.
►
Rinse lumens with an appropriate disposable
syringe at least five times.
►
Drain any remaining water fully.
Phase V
►
Dry the product in the drying phase with suitable
equipment (e.g. cloth, compressed air), see Vali-
dated cleaning and disinfection procedure.