B. Braun Aesculap Spine activ C FW857R Gebrauchsanweisung Seite 10

Einsetzinstrument
Vorschau ausblenden Andere Handbücher für Aesculap Spine activ C FW857R:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 18
Aesculap
®
activ C-Einsetzinstrument
Validated cleaning and disinfection procedure
Validated procedure
Manual cleaning with
immersion disinfection
Manual pre-cleaning
with brush and subse-
quent mechanical alka-
line cleaning and ther-
mal disinfection
8
Specific requirements
Cleaning brush: 250 mm/
∅ 3.7 mm, e.g. GK469200
20 ml disposable syringe
When cleaning instruments with
movable hinges, ensure that these
are in an open position and, if
applicable, move the joint while
cleaning.
Drying phase: Use a lint-free cloth
or medical compressed air
Cleaning brush: 250 mm/
∅ 3.7 mm, e.g. GK469200
20 ml disposable syringe
When cleaning instruments with
movable hinges, ensure that these
are in an open position and, if
applicable, move the joint while
cleaning.
Place the instrument in a tray
that is suitable for cleaning
(avoiding rinsing blind spots).
Connect components with lumens
and channels directly to the rins-
ing port of the injector carriage.
Place instruments in the tray with
their hinges open.
Reference
Chapter Manual cleaning/disinfection and
sub-chapter:
Chapter Manual cleaning with immer-
sion disinfection
Chapter Mechanical cleaning/disinfection
with manual pre-cleaning and sub-chap-
ter:
Chapter Manual pre-cleaning with a
brush
Chapter Mechanical alkaline cleaning
and thermal disinfecting