Problem
Possible Cause
Alarm Sound
Incorrect Power-Off Sequence.
Continuous, No
Lights
No power, battery drained.
Internal part failure.
Alarm sounds weak.
Internal part failure.
Low battery
Loose lead
Connections between
Light remains on.
electrodes lead wires, and
patient cable are not properly
made.
Defective lead wires.
Defective patient cable.
Defective electrodes.
Instructions
Press the POWER button and insure that the power light
is illuminated.
Press and hold the RESET button. Press and release
POWER. Wait two seconds then release the RESET
button.
Connect power cord/battery charger. Use Power-Off to silence
alarm.
Press and hold the blue RESET button.
Press and release the gray POWER button. Wait 2
seconds, then release the RESET button.
Prior to use, allow battery to charge approximately 30 minutes.
You may then operate the monitor while it is plugged in. Allow
the battery to charge for 6 hours before using the monitor on
battery power.
Contact your dealer.
Contact your dealer.
Charge battery.
Verify that
(a) baby's skin underneath the electrodes is clean,
(b) electrodes are clean, and
(c) lead wires are fully inserted into the electrodes and patient
cable.
Replace lead wires and perform Functional Self-Test.
Replace patient cable and perform Functional Self-Test.
Replace electrodes.