Turning the SmartMonitor 2 on

You have properly set up your SmartMonitor 2, understand how the monitor functions and how to respond to alarms. You are
now ready to begin monitoring your baby's breathing and heart activity according to the schedule prescribed by your health
care professional.
Push the POWER button. The SmartMonitor 2 performs a system check. The lights on the front of the SmartMonitor 2 and the
alarm come on briefly. Within 10 seconds, the green respiration, and heart lights begin to blink. If the lights do not blink,
check that you have attached the electrode belt properly to the baby, that the lead wires are pushed in, and that the cables
are connected.
Once your baby is properly connected to the SmartMonitor 2 and the power is on:
The green (battery) charger light is on.
The green power light is on.
The green respiration light and the green heart light are blinking.
All other lights should be off.
If the lights do not blink, refer to the steps found in the Setting up the SmartMonitor 2 section in this manual and be sure
you have followed all instructions.

Turning the Monitor off - Sibling Alarm

The SmartMonitor 2 has a built-in safety feature called a sibling alarm. If the monitor is not turned off in a specific sequence,
the green power light will remain on and the alarm will sound continually. This safety feature makes sure the power is not
accidentally turned off. To turn the SmartMonitor 2 off:
Press and hold the blue RESET button.
Press and release the gray POWER button.
Wait 2 seconds then release the RESET button.