Herunterladen Inhalt Inhalt Diese Seite drucken

Modem Auto Answer In Communications Mode - Circadiance SmartMonitor2 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 324
If there is an emergency and access to the telephone is
required while modem is in use, remove phone cord
from modem slot and use a working phone.
Plug the telephone wire into the modem connector on the
back of the SmartMonitor 2 (see illustration).
Plug the other end of the phone wire into the wall phone
jack.
The SmartMonitor 2 will continue to monitor your baby
during download. All alarms are functional.
The monitor will continue calling every 30 minutes, for 4 hours, until it connects with the computer. If no connection is
made, the monitor will try again the next day at the pre-selected time. For example, when the modem is programmed
to call at 7am but fails to make a connection after 8 attempts it will stop trying. The next day at 7am it will try again.
This will continue until a connection is made.

Modem Auto Answer in Communications Mode

1. Connect a phone cord from the wall jack to the modem connector on the back of the SmartMonitor 2.
2. To allow the SmartMonitor 2 to work with the modem, place the monitor in Communications Mode:
Press and hold the blue RESET button.
Press and release the POWER button.
Wait until the monitor alarms. Now release the RESET button.
Briefly press and release the RESET button again.
The SmartMonitor 2 beeps every 10 seconds whenever it is in the Communications Mode. This is a reminder that the
SmartMonitor 2 is powered on for working with the computer, memory card, or modem, but NOT for monitoring the
infant.
Do not connect the infant to the SmartMonitor 2 when in the Communications Mode, the apnea and heart rate
alarms are not operational in this mode.
Transferring SmartMonitor Information