Performing a Functional Self-Test

The SmartMonitor 2 functional self-test checks that all the features of the unit are functioning properly. You should perform a
functional self-test at least once a week or according to the instructions given by your health care professional. You should
also perform the test:
After a lead wire is changed
After the patient cable is changed
To perform the functional self-test, follow the steps listed below.
Insert the patient cable into the socket located on the front of
the monitor.
Connect the lead wires to the patient cable. Inset the white
lead wire into the opening labeled RA. Insert the black lead
wire into the opening labeled LA.
Connect the lead wires to the functional self-test socket on
the side panel of the SmartMonitor 2. Insert the white lead wire
into the RA an opening and then the black lead wire into the LA
opening.
Turn on the monitor. You will hear a short beep and the lights
on the front come on briefly then go off.
After all the alarm lights go out, the green power and charger lights remain on and the green heart and respiration lights
are blinking.
The heart and respiration lights continue to blink for about 30 seconds.
When the green lights stop blinking, the red low heart light will come on within about 7 seconds and the alarm beeps once
every second.
Next, the red apnea light comes on (the amount of time before the red apnea light comes on is determined by the Apnea
Delay parameter selected at the time the monitor was set-up) and the low (heart) light remains on. (There should be no
green heart or respiration light flashes during this time).
Follow the instructions in the Self-Test Troubleshooting section, if necessary.
Remove the lead wires from the functional self-test socket.