Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

To prevent permanent damage to this unit, the unit must reach room temperature prior to conducting any setup
and/or unit operations.
It is important to ensure that the bed is working properly before it is put into service. The following list will help
ensure that each part of the bed is checked.
• Install the foot and headboards on the bed . Insert the footboard carefully so that the footboard fit in smoothly .
CAUTION
Because individual beds may have different options, footboards should not be moved from one bed to another . Mixing
footboards could result in unpredictable bed operation .
• Plug the power cord to the bed connector at the head end of the bed and into a properly grounded hospital grade
wall socket .
• Turn the battery power switch to the "ON" (1) position (located below the litter surface at patient right side) .
• Set the time and date through the touch screen (see
WARNING
This bed is equipped with a hospital grade plug for protection against shock hazard . It must be plugged directly into a
properly grounded receptacle . Grounding reliability can be achieved only when a hospital grade receptacle is used .
• To test the manual backup brake (located on the patients right side only), flip the manual backup brake pedal
outward and depress down fully on the foot end side of the pedal . The word BRAKE in red will be visible in the
clear window located on the base next to the pedal and the Brake symbol will be flashing on the footboard control
panel . The brakes should now be applied .
• To test the electric brake; engage the brake by pressing the brake button on the siderail or footboard control panel .
Try moving the bed to ensure the brakes are applied . Press the Neutral or Brake button on the siderail control panel
or on the footboard to disengage the brake .
• For Model 2152/2154/2156 Beds only: On the patient's right side of the bed, flip the manual backup brake pedal
outward and depress down fully on the head side of the pedal . The letter D in GREEN should be visible in the clear
window located on the base next to the pedal . The Zoom
fully on the center of the pedal until the letter N in BLUE is visible in the clear window . The Zoom
should now be disengaged .
• Ensure that the siderails raise, lock in the fully raised position and lower smoothly .
• Run through each control on the footboard .
• Verify the scale system and the Bed Exit system for proper operation .
• Run through each control on both inner and outer control panels of the head siderails . If the bed is equipped with
the optional Communication Pendant, plug the provided cable to the 37-pin connector located at the head end of
the bed and into the proper wall socket .
• Raise the bed completely and activate the Trendelenburg function . Ensure the head end lowers to the full down
position . See "Nurse Control" illustration on
• Raise the bed completely and activate the reverse Trendelenburg function . Ensure the foot end lowers to the full
down position . See "Nurse Control" illustration on
• Verify functionality of the CPR emergency release: raise the fowler up then depress down on the CPR pedal
located on either side of the bed at the head end of the base . The fowler will lower towards a flat position until
the pedal is released .
• Ensure the Knee Gatch (if raised) also starts flattening when the Fowler is completely down . Following the
complete lowering of the Fowler, wait approximately 30 seconds (the time for the Fowler control motor to reset)
and verify that the motor has indeed reset by raising the Fowler fully up using the "Fowler Up" control .
• Verify the following optional equipment for proper operation: 120V auxiliary outlet, night light, emergency crank
operation, etc .
www.stryker.com

Setup Procedures

page 1-39
for procedure) .
®
Drive wheel should now be engaged . Depress down
page
1-22 . Level the bed using the Bed Lift system controls .
page
1-22 . Level the bed using the Bed Lift system controls .
2151-109-005 REV A
Drive wheel
®
Return To Table of Contents
1-13

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis