OPTIONAL ZOOM

®

DRIVE SYSTEM (MODEL 2152/2154/2156 ONLY)

If your bed is equipped with the Optional Zoom
continue with the procedures listed below .
• With the battery power switch in the "ON" position and the drive wheel in the neutral position (not touching the
floor), ensure the "Engage Drive Wheel" LED on the head end control panel is illuminated .
• Run through the operation of the drive wheel (refer to
• On both sides of the bed, flip the manual backup brake pedal outward and depress down fully on the head side of
the pedal . The letter D in GREEN should be visible in the clear window located on the base next to the pedal . The
Zoom
®
drive (2152/2154/2156 Model only) should now be engaged . Depress down fully on the center of the pedal
until the letter N in BLUE is visible in the clear window . The Zoom
www.stryker.com
Setup Procedures
®
Drive System, run through the preceding setup procedures and
page
1-54) to ensure it is operating properly .
®
2151-109-005 REV A
drive should now be disengaged .