It is important that the Model 1025 Stretcher is working properly before it is put into service . The following list will help
ensure that each part of the unit is checked .
• Plug the power cord into a properly grounded, hospital grade wall receptacle . The 12 volt batteries that provide power
to the drive wheel and backup power to the unit functions will charge whenever the power cord is plugged into the
power source . The batteries require approximately 6 hours of charging time before the stretcher is put into service
• Depress the pedal at either end of the stretcher fully to set the four wheel brakes and verify all four casters are
locked .
• Ensure the siderails raise and lower smoothly and lock securely in the full up position .
• Run through the operation of the drive wheel (see "Transporting the Stretcher Using the Drive Wheel" on
and "Operating the Glideaway Siderails" on
CAUTION
The Model 1025 Stretcher is equipped with a hospital grade plug for protection against electric shock hazard . It must
be plugged directly into a properly grounded three-prong receptacle . Grounding reliability can be achieved only when a
hospital grade receptacle is used .
www.stryker.com

Setup Procedures

page
1-15) to ensure it is operating properly
1025-009-005 REV B
page 1-13
Return To Table of Contents
1-9