Renishaw SP25M Bedienungsanleitung Seite 53

Vorschau ausblenden Andere Handbücher für SP25M:
Inhaltsverzeichnis

Werbung

SP25M user's guide
http://www.renishaw.com
Test setup criteria
Test site
Renishaw UK
Styli used
Renishaw's M3 stylus range
3
CMM
U
 = 0.48 µm + L / 1000
specification
CMM
Renishaw's UCC2
controller
Probe
Renishaw's third order polynomial calibration method
calibration
Data filter
Harmonic simple cut off order = 60 UPR (undulations per revolution)
used
Artefacts
Ø25 mm (nom) calibration sphere (for the ISO 10360 Part 2 / Part 4 tests)
used
Ø50 mm (nom) ring gauge (for the ring gauge scan tests)
Scanning
5 mm/s (for the ISO 10360 Part 4 and ring gauge scan tests)
speed
Touch
5 mm/s (for the ISO 10360 Part 2 tests)
speed
Back off
1 mm/s (for the ISO 10360 Part 2 tests)
speed
Bi-
The effect of probe misalignment and contact friction will result in radius changes with scanning direction if a suitable
directional
calibration procedure is not used.
scans
Mechanical hysteresis within a probe mechanism is best demonstrated by a bi-directional scan.
Therefore, Renishaw have decided to take the high integrity approach of quoting ring gauge metrology using bi-directional
scan data to demonstrate the performance of the Renishaw probe and calibration method.
Issued 10 2015
53